PLL Testing Using the Femto Continuous Time Interval Analyzer

Introduction:

Integrated circuit testing is becoming ever more complicated with increasing clock rates, lower timing margins and the proliferation of on-board phase locked loops (PLL's). With this in mind, Guide Technology has introduced the Femto Continuous Time Interval Analyzer (CTIA), a new concept in time measurement instrumentation. Much faster than previously available time measurement instrumentation, the Femto features truly parallel measurements, rapid time-correlated measurement rates, and multiple Digital Signal Processor (DSP) processing power. The Femto can test a wide variety of PLL circuits with a maximum frequency of over 800 MHz. PLL measurements such as frequency and jitter can typically be accomplished in 100 milliseconds for 64 points. The Femto frequency measurement features typical accuracy of 1 part per million (ppm). The Femto can also make extremely precise jitter measurements with a jitter noise floor of +/- 3 picoseconds. NOTE: The programming details in this application note apply to the Femto 2000. Please refer to the User Manual for Femto 3200 implementation.

What Is the Femto ? The Femto is a true Continuous Time Interval Analyzer (CTIA), unlike many timing instruments which are merely frequency or time counters. The Femto can be used in either stand-alone mode or integrated into a production test environment. Front Panel Operation The Femto is accurate enough to be used as a high precision lab instrument. Operation in lab instrument mode is from the front panel using a mouse and keyboard for user input. The Femto includes a large display screen on its front panel for easy data and graphics display. Production Oriented Design/Operation The Femto design is optimized for high-throughput testing in an automated test equipment (ATE) environment. The Femto can take measurements on up to 8 channels simultaneously. Rapid and powerful measurement calculations are processed by up to 4 on-board digital signal processors (DSP's). Control of the instrument in production mode is via a GPIB interface.

Configuration:

With A Tester The Femto has been configured to work with virtually any ATE system currently in use. Instrument control and data transfer is via GPIB interface. Programming is accomplished using the industry standard "C" programming language and custom, high level device driver calls. Femto driver code is available for both Unix and Windows platforms. Typical ATE Setup Loadboard Interface A key part of any ATE application is the loadboard signal interface by which the Device Under Test (DUT) signals to be sent to the Femto . Guide Technology provides several choices for loadboard signal interfaces. The loadboard interface can be via a pogo block connector mounted in the testhead, GT QuikConnect connectors mounted on the loadboard and testhead or traditional SMA connectors mounted on the loadboard. The signals are sent from the loadboard interface connectors to the Femto via high-bandwidth coax cables. PLL Measurements In Production Definition of Phase-Locked Loop (PLL) A phase-locked loop (PLL) is a circuit that synchronizes the phase of a locally generated signal with that of an external signal in any electronic system. Examples of PLL applications 1. Skew & clock delay control in large digital systems 2. Frequency shift correction in satellite transmissions 3. Local & reference oscillator locking in analog video equipment.

A phase detector is used to determine the phase difference between an external reference signal (Fextern) and a local signal (Flocal). The Flocal signal is generated by an on-board Voltage Controlled Oscillator (VCO). Since the phase detector circuit requires that the two comparison signals be of the same frequency, a "divide-by-N" circuit is inserted between the VCO output (Flocal) and the phase detect circuit so that the two signals are at the same frequency for proper phase comparison in the phase detector. The voltage produced by the phase detector (Vph_diff) represents the difference in phase between the two signals being phase-locked. This phase difference voltage (Vph_diff) is filtered by a Loop Filter to remove any high frequency noise (ripple) and then fed into the voltage controlled oscillator (VCO) - the same VCO which is used to generate the Flocal signal and which adjusts Flocal to match the phase of Fextern using the Vph_diff signal. Common PLL Measurements The two most common PLL measurements tested are Average Frequency and Timing Jitter.

Average Frequency The Average Frequency Measurement determines how close the PLL output frequency is to the desired frequency. Several different types of frequency measurements are possible with various trade-offs of accuracy and test time. Frequency measurements are discussed in more detail later in this application note. Jitter Timing Jitter is a measurement of how much the PLL output frequency or output period varies from the desired value. Of particular concern are PLL's in digital systems which face constraints on the variation allowed in the length of individual output periods. RMS Jitter Cycle jitter is determined by taking many measurements of the PLL output period. The Femto on-board DSP's can quickly calculate standard deviation of these output single period measurements. The standard deviation value is equal to the RMS jitter of the output clock. If the distribution of the single period measurements is Gaussian then the cycle jitter is considered random and the RMS jitter can be used to project the range of the jitter present. Peak to Peak Jitter Peak to Peak Jitter is determined by the range of the single period measurements in the measurement block. The Femto on-board DSP's can also quickly calculate the minimum and maximum periods that were measured. The Peak to Peak jitter measurement is the difference of the maximum and minimum measured periods. Peak to Peak Jitter measurement is described in more detail later in this application note. RMS Jitter vs. Peak to Peak Jitter The Peak to Peak jitter measurement is an actual measurement of the range of the single periods present while the RMS jitter is used to project the range of the single periods that will be present.

The advantage of the RMS jitter measurement is that the RMS value can be established after a comparatively few single period measurements (typically 50 or more) while Peak to Peak jitter measurement will continue to increase as more and more periods are measured (stabilizing somewhat after perhaps thousands of single period measurements). Long Term Jitter The Long Term Jitter measures cycle jitter as well as low speed frequency variations (frequency drift) that will not be show in the previously described cycle jitter measurements. Long Term jitter uses a special Femto mode called Time Interval Error (TIE) Mode. The Long Term Jitter measurement is described in more detail later in this application note. How Does Femto Make Timing Measurements? An extensive description of the Femto many modes and features is beyond the scope of this application note. However, some knowledge of the Femto operation is helpful. The Femto is capable of many types of measurements. Continuous, Time-correlated Timing Measurements The Femto CTIA makes direct measurements of event timing triggered by user-defined edge polarity and threshold voltage. Every edge measured by the Femto CTIA is "time stamped" with two pieces of information: the time Tn and the event En relative to the time stamp (T0, E0) of the first edge measured in a set of measurements. That is, every edge timing measured in a signal stream is correlated in time relative to the first edge timing measured, T0, in a set of measurements. Every event measured in a signal stream is correlated in time relative to the first event counted, E0, in a set of measurements The easiest type of measurement to understand is the two-edge measurement such as Single Period or Pulse Width. The Femto takes the specified measurement, records the measurement, records the time at which the measurement occurred and records the event count of the measurement. This constitutes a single measurement. An example of a single edge measurement is the frequency measurement. The Femto records the time at which the edge occurred and records the event count of the measurement. The frequency measurement is calculated by subtracting the time of the previous edge from the time of the current edge and dividing by the number of edges observed. Usually a large group of measurements are taken at one time by repeating either of the preceding processes. The statistics of the measurement group are then be calculated by the Femto on-board DSP processors.

These statistics include the average measurement value, the minimum measurement value, the maximum measurement value and the standard deviation of the measurement values. Arming Arming controls when the each measurement actually occurs. The Femto supports a wide variety of timing modes including external arm (take a measurement when an external signal occurs), arm by time (take measurements at a specified rate) and arm by time after an external arm. Voltage Threshold & Polarity A voltage threshold & edge polarity must be specified in order for the Femto to make measurements. A measurement will occur when the input edge crosses the specified voltage threshold. The voltage threshold can be set as either a fixed voltage or as a percentage of the input signal's voltage swing. Statistical vs. Time Domain Measurements Two techniques are commonly used to make relevant timing measurements from the block of measurements gathered by the Femto . The most common technique is to derive the basic statistics of the measurement block. The Femto DSP's automatically return the following statistics for any measurement block: MEAN (average), MIN (minimum), MAX (maximum) and STD DEV (standard deviation). A second technique is for the entire block of measurements to be used instead of just the standard statistics. The Femto automatically records the time at which each of the measurements were taken. Plotting the measurement values vs. the time when they were taken allows analysis to be done in the time domain. The following is a comparison of statistical and time-domain test techniques: PLL Statistical Measurements The Femto typically measures a timing parameter by first making a large group of measurements. The user then has the option of having the statistics calculated for that group of measurements or having all of the measurements returned to the tester. If just the statistics are needed, the of measurements are processed by the on-board DSP's. The statistics of the group are returned to the tester. The DSP-derived statistics automatically calculated and returned to the tester are the mean (average) of the group of measurements, the minimum measurement of the group, the maximum measurement of the group, and the standard deviation (RMS) of the group. The tester then makes a pass/fail decision based on the statistics.

Statistical Analysis Summary:

Step 1: Femto takes a block of time measurements from the PLL output.

Step 2: Femto calculates the timing statistics from the block of timing measurements (mean, maximum, minimum, standard deviation).

Step 3: Timing statistics are transferred to the tester.

Step 4: Tester datalogs and makes pass/fail decisions based on the timing statistics.

Advantages:

Faster test time because only the statistics are passed to the tester. Disadvantages: Pass/fail limited to a few standard statistical measurements. Timing information may be obscured by statistical process. Time domain tests not possible. PLL Time-Domain Measurements The entire group of timing measurements made by the Femto can be returned to the tester if the standard statistical analysis is not sufficient. The tester can then make pass/fail decisions based on time-domain or other analysis of the raw time measurements. The Femto can also return the time that each measurement was made along with the measurement value. Knowing the value of each measurement and when the measurement was made allows analysis in the time-domain. The Femto- high measurement rate of up to 1 Megasample per second allows direct, single-shot measurement and analysis of PLL performance in the time domain.

Much of this type of time domain analysis may be missed using averaging techniques and statistical techniques common to many other instruments. Measurements that can be made only in one-shot mode include repeated variations of the PLL frequency and problems in the lock acquisition and loop performance of the PLL. Time domain analysis can be useful in production test programs where measurements are required which cannot be done with statistical techniques. Time domain analysis can also be useful for characterization test programs where design criteria and device performance are verified to an extent not required for production. The Femto can also be used as a bench-type instrument using the front panel controls and the front panel display to perform time domain analysis in a debug mode.

Time-Domain Analysis Summary:

Step 1: Femto takes a block of time measurements from the PLL output.

Step 2: The block of timing measurements are transferred to the tester.

Step 3: The time domain measurements are calculated from the block of data.

Step 4: Tester datalogs and makes pass/fail decisions based on the time domain measurements.

Advantages:

All timing information available to the tester. Time domain tests are possible. Disadvantages: Slightly longer test time due to more data being passed. PLL Measurement Details Introduction We can now review the details of some of the more common PLL test measurements being made today. Details for the following measurements are included in this application note: Statistical tests: Average Frequency Measurement Single Period Measurement Frequency Measurement Using Single Period RMS Jitter Measurement Peak to Peak Jitter Measurement: Deterministic Jitter Long Term Jitter (Frequency Wander) Pulse Width Measurement Duty Cycle Calculation Rise Time / Fall Time Measurements Voltage Related Tests: Output Level Searches Time Domain Tests: PLL Lock Time (PLL Settling Time) PLL Loop Filter Bandwidth PLL Frequency Modulation Test Time Reduction Technique: Multiple Block Measurement Mode Average Frequency Measurement: Measuring the Average Frequency is the most basic PLL test. The FREQUENCY AVERAGE measurement mode provides the most accurate means of measuring frequency.

First, a time interval is selected (3 us for best accuracy). The Femto automatically counts number of signal edges during each time interval. The frequency is then calculated for each time interval. The Average Frequency and other statistics of the group of measurements are returned to the tester.

Step 1: Set up the DUT so that the PLL output is clocking and stable.

Step 2: Take the PLL output frequency measurements. MODE = FREQUENCY AVERAGE MEASUREMENT ARM = BY TIME, 1us or greater interval. SAMPLE COUNT = User Specified Figure 3 - Frequency Average Measurement Mode

Step 3: Read back the Frequency Average statistics and make PASS/FAIL decision based on the MEAN value. Single Period Measurement: The SINGLE PERIOD Measurement mode provides measurements of the time delays of individual periods in the PLL output. The single period results can be read back either as the basic statistics (mean, minimum, maximum and standard deviation) or else the single period measurements can all be read back in a block read. The single period measurement statistics provide enough information for several common tests, as shown in the following examples. Each individual period measurement is effectively an instantaneous frequency measurement. Plotting the single period measurements vs. time allows time-domain analysis of the PLL output frequency.

Step 1: Set up the DUT so that the PLL output is clocking and stable.

Step 2: Take the single period measurements. MODE = SINGLE PERIOD MEASUREMENT ARM = BY TIME, 1us or greater interval. SAMPLE COUNT = User Specified Figure 4 - Single Period Measurement Mode & Femto- Display Step 3: Read back the Single Period statistics and make PASS/FAIL decision based on the MEAN, MIN amd MAX values. Frequency Measurement Using Single Period: The average frequency can be calculated from the Single Period Measurement statistics by inverting the MEAN (average) period measurement. Test time can be saved by this method since the FREQUENCY AVERAGE test can be skipped.

Step 1: Obtain the Single Period MEAN statistic from the Single Period test..

Step 2: Calculate the Frequency Average MEAN statistic and make PASS/FAIL decision based on test limits. RMS Jitter Measurement: The RMS Jitter of a signal can be calculated by taking the standard deviation of the Single Period measurement above. Knowing the RMS jitter is useful because allows calculation of other jitter statistics if the jitter follows a Gaussian distribution. Step 1: Obtain the Single Period STD DEV statistic from the Single Period test. Step 2: Make PASS/FAIL decision based on the Single Period STD DEV and the test limits.

Peak to Peak Jitter Measurement:

The Peak to Peak Jitter of a signal is the total spread of the single period measurements. The Peak to Peak Jitter is the measured "worst case" jitter. The Peak to Peak jitter can be calculated by the computing the range of measurements made in the Single Period measurement above. The range is the Maximum Single Period measurement minus the Minimum Single Period measurement. Note that the Peak to Peak Jitter measurement will tend to grow as the sample count grows while the RMS Jitter will tend to stabilize while the sample count grows.

Step 1: Obtain the Single Period MINIMUM and Single Period MAXIMUM statistics from the Single Period measurement above.

Step 2: Calculate the Peak to Peak Jitter using the following equation: Peak to Peak Jitter = Single Period MAX - Single Period MIN

Step 3: Make PASS/FAIL decision based on the Peak to Peak Jitter and the test limits. Deterministic Jitter Deterministic Jitter is timing jitter that is not random. Many times Deterministic Jitter is inherent in the design of a particular PLL. Deterministic Jitter is often periodic. The PLL design displayed in Figure 5 has periodic Deterministic Jitter due to the PLL divider stages. Figure 5 is an example of PLL analysis in the time domain that would not be possible using statistical or oscilloscope techniques.

Long Term Jitter is the deviation of the PLL output from an "ideal" clock output over a long period of time. While Long Term Jitter measures the variations in individual cycle periods, as does a Single Period Jitter measurement, a Long Term Jitter measurement also measures longer term variations in frequency, also known as frequency wander. The Femto provides a special measurement mode called Time Interval Error (TIE) to measure Long Term Jitter. The Time Interval Error mode reports the time difference between the measured PLL output edges and an "ideal" edge calculated from a specified frequency. The TIE mode is used to measure variation or drift from the ideal clock frequency and phase as well as to measure short-term cycle jitter. The TIE results will include the effects of both short and long term timing error in the PLL output. TIE results can be read back either as statistics (mean, min, max, std dev) or the individual timing error measurements can all be read back in a block read.

TIE results can be affected by inaccuracy in the "ideal" frequency specified. The "ideal" frequency is usually the result from a previous frequency measurement. Care must be taken that the "Ideal" frequency value used is very accurate or the TIE errors will grow continuously due to the mismatch of the "ideal" PLL frequency and the actual PLL frequency.

Step 1: Determine the PLL output "ideal" frequency by using previous frequency measurements.

Step 2: Set up the DUT so that the PLL output is clocking and stable.

Step 3: Take the TIE measurements. MODE = Time Interval Error (TIE) SAMPLE COUNT = User Specified MEASUREMENT ARM = BY TIME, 1us or greater interval Figure 6 - Time Interval Error (TIE) Measurement Mode

Step 4: Make PASS/FAIL decision based on the Minimum TIE and Maximum TIE statisitcs and the test limits. Example: Frequency Wander The TIE measurement selects certain edges of an input signal, records the transition times of those edges, and subtracts from those times the corresponding transition times of an ideal signal of the same frequency. The resulting series of numbers, called time interval errors, or TIEs, indicate how far each of the measured edges is from its ideal position. A positive TIE indicates that the corresponding input signal transition lags its ideal position.

A negative TIE indicates that the corresponding input signal transition leads its ideal position. The minimum and maximum TIE measurements represent the amount of long-term jitter (wander) in the PLL output. Pulse Width Measurement: The PULSE WIDTH Measurement mode measures either the width of the positive going pulse (Pulse Width High) or the width of the negative going pulse (Pulse Width Low). The PULSE WIDTH Measurement mode provides a block of measurements, as did the Single Period test earlier. Reading back the Femto DSP generated statistics can provide the average pulse width (mean), the smallest pulse width measured (minimum), the largest pulse width measured (maximum) and the pulse width jitter (standard deviation). As with other tests, the pulse width results can all be read back in a block read also. Unlike the Frequency and Single Periods measurements,

Pulse Width measurements can be influenced strongly by the voltage threshold setting and also by signal distortion. This is because the Pulse Width test measures edges of opposite polarities (rise to fall or fall to rise) so that pulse slew rates and distortion can add or subtract from the reading. Frequency and Single Pulse measurements are made on edges of the same polarity (rise to rise or fall to fall) as so slew rates and signal distortion on one edge tend to offset the same problems on the second edge.

Pulse Width High Measurement:

Step 1: Set up the DUT so that the PLL output is clocking and stable.

Step 2: Take the Pulse Width High measurements. MODE = PULSE WIDTH HIGH MEASUREMENT ARM = BY TIME, 1us or greater interval. SAMPLE COUNT = User Specified THRESHOLD = FIXED or PERCENTAGE

Step 3: Read back the Single Period statistics and make PASS/FAIL decision based on the MEAN, MIN, MAX and STD DEV values. Figure 7 - Pulse Width Measurement Mode Pulse Width Low Measurement:

Step 1: Repeat the Pulse Width High test using Pulse Width Low mode. MODE = PULSE WIDTH LOW Duty Cycle Calculation: The clock Duty Cycle is the ratio of the Mean Pulse Width High to the Mean Single Period as in Equation 1 below. Like Pulse Width measurements, Duty Cycle measurements can be affected by threshold settings and signal distortion.

Step 1: Calculate Duty Cycle using the following equation: Equation 1: Calculate the PLL Duty Cycle using the pulse width high and single period measurements. MEAN of Pulse Width High Duty Cycle = --------------------------------------- MEAN of Single Period

Step 2: Make PASS/FAIL decision based on the calculated Duty Cycle value. Rise Time / Fall Time Measurements: The RISE TIME Measurement mode provides measurements of the rise time delay between two different voltage thresholds on the same rising signal edge. The FALL TIME Measurement mode does the same for the fall time delay between two different voltage thresholds on the same falling signal edge. Typically only the MEAN rise and fall time statistics are used for test purposes. Rise Time and Fall Time measurements use both a lower and an upper voltage threshold setting. These two voltage threshold settings can be programmed either as fixed voltage levels or as percentages of the overall voltage swing. Rise Time and Fall Time measurements can be influenced strongly by incorrect voltage threshold settings and also by signal distortion.

Step 1: Set up the DUT so that the PLL output is clocking and stable.

Step 2: Take the Rise Time (or Fall Time) measurements. MODE = RISE TIME (or FALL TIME) MEASUREMENT ARM = BY TIME, 1us or greater interval SAMPLE COUNT = User Specified LOWER THRESHOLD = FIXED or PERCENTAGE UPPER THRESHOLD = FIXED or PERCENTAGE

Step 3: Make PASS/FAIL decision based on the Mean Rise Time measurement (or Mean Fall Time measurement) and the test limits. Risetime Measurement Mode Output Level Searches The Femto provides a high-speed, automated routine to determine the maximum and minimum voltages found in an input waveform. The results of the voltage searches can be used to test the PLL output VOL and VOH levels. The search results can also be used to set voltage thresholds to a percentage of the voltage swing. The voltage search routine typically executes in 100 milliseconds or less for 8 channels.

PLL Lock Time (PLL Settling Time): The time it takes for a PLL to adjust to a new frequency setting is called the PLL "lock time" or "settling time". The PLL lock time can be critical in many applications where changes to the PLL output frequency must be stabilized quickly. Important PLL performance information can be obtained by watching the PLL response as it acquires lock. The Femto can perform two different kinds of lock time tests. A "go/nogo" lock time test simply tests whether lock has been acquired at the specified time. A "time-to-lock" test measures the actual lock time for the PLL. The "time-to-lock" measurement test provides all of the single period measurements as the PLL adjusts and locks. In a debug environment, the Femto single period measurements can be examined graphically from the front panel or can be read and processed as a block of measurements. PLL Lock Time ("Go/NoGo" Test Method): The PLL "Go / NoGo" Lock Time test simply measures the PLL output frequency after the specified lock time. The device passes if the measured frequency is in the specified frequency range.

Step 1: Set up the DUT so that the PLL output is in its initial condition.

Step 2: Setup the Femto to take the measurements MODE = FREQUENCY AVERAGE MEASUREMENT ARM = BY TIME, 1us interval BLOCK ARM = EXTERNAL SAMPLE COUNT = User Specified

Step 3: Begin the PLL Lock sequence by programming the programming the PLL or changing its input frequency. An EXTERNAL ARM is sent after the specified PLL settling time. The External Arm signals the Femto to begin measuring the PLL output frequency. Figure 9 - External Arming of Frequency Average Measurement Step 4: Verify that the MEAN statistics value of the FREQUENCY AVERAGE measurement is within the test limits for the PLL output frequency. PLL Lock Time ("Time-to-Lock" Measurement Method): The PLL Lock Time "time-to-lock" measurement test finds the actual time when the PLL acquired lock. This test is performed with the Femto in single period measurement mode. As mentioned previously, each single period measurement is actually an instantaneous frequency measurement. The block of data obtained by the test program is the actual instantaneous frequency measurements as the PLL acquired lock. Measurements are made at the maximum rate of 1 MHz (1 us intervals).

Step 1: Set up the DUT so that the PLL output is in its initial condition.

Step 2: Setup the Femto to take the measurements MODE = SINGLE PERIOD MEASUREMENT ARM = BY TIME, 1us interval BLOCK ARM = EXTERNAL SAMPLE COUNT = User Specified Figure 10 - PLL "TIME-to-Lock" Measurement

Step 3: The blocks of single-period data and associated elapsed-time data are read back to the tester after the measurement has been completed.

Step 4: The test program calculates the lock time by examining the data block and determining when the measured frequency entered the specified lock range.

Step 5: If desired, the PLL frequency can be plotted against the elapsed time to provide a complete picture of the PLL lock response. PLL Loop Filter Bandwidth: The speed with which a PLL reacts to frequency and phase errors is called the PLL Loop response. The PLL Loop Response is determined largely by the PLL Loop Bandwidth. The PLL Loop Bandwidth is therefore an important parameter in determining PLL performance. Too little bandwidth results in a PLL that is slow to lock to a new frequency. Too much bandwidth will result in a PLL which is quick to move to a new frequency but which may tend to oscillate, may over-shoot the desired frequency and may pass input jitter to the output. Standard statistical analysis of the PLL frequency and jitter cannot be used to measure the PLL loop performance. Statistical analysis of a PLL with poor loop performance may indicate a poor jitter performance but there is no way to use statistical analysis to determine the PLL loop performance. Time domain analysis can be used to not only measure the loop bandwidth but to also graph the loop response, including over-shoot and under-shoot. PLL loop bandwidth is tested by applying a phase shift to the input signal of the PLL. The phase shift can often take the form of a skipped input clock. The phase shift of the input clock causes the PLL to adjust (phase-lock) to the new input phase. The speed at which the PLL phase-locks to the new input phase is a measure of the PLL's loop response and loop bandwidth. Time Interval Error (TIE) mode is used in this measurement. In TIE mode, change in the PLL output phase is measured as a phase error relative to an "ideal" output clock.

Step 1: Determine the PLL output "ideal" frequency by using previous frequency measurements.

Step 2: Set up the test vectors to provide an external arm after the PLL output is clocking and stable.

Step 3: Set the test vectors to provide a input phase shift or skipped input clock shortly after the external arm.

Step 4: Take the TIE measurements. MODE = Time Interval Error (TIE) SAMPLE COUNT = User Specified MEASUREMENT ARM = BY TIME, 1us or greater interval BLOCK ARM = External Arm Step 5: Continue measurements until the PLL output has completed its phase adjustment. Figure 11 - Example of phase shift applied to PLL input signal Step 6: Calculate the total bandwidth of the PLL output using the following equation: PLL Loop Bandwidth = 0.35 / risetime where risetime is the time it took for the PLL to settle to the value of the induced phase shift error from the time the input signal was phase shifted (e.g. 50ns). Figure 12 - Graph of Time Interval Error (TIE) measurements indicating time for PLL to lock onto applied input phase shift PLL Frequency Modulation: PLL Frequency Modulation occurs when the PLL output frequency varies over time. PLL modulation can be deliberate, such as spread spectrum applications, or it can be accidental, which results in poor jitter performance. PLL Modulation is difficult to measure statistically beyond simply finding that the PLL output has excessive jitter. The Femto front panel display can often display PLL modulation in the time domain.

The statistical information includes a large standard deviation (jitter) but there is no indication as to the cause of the jitter. The histogram is more informative than the raw statistic numbers but again it gives no indication of why the measured frequency range is so large. The upper left diagram (red) of Figure 13 plots the time domain information of PLL output frequency versus time. The effect of modulation is clearly shown as the measured frequency rises and falls over time. This type of modulation is not observable when the timing information is reduced to the standard statistics.

The effect of modulation is clearly seen as the measured frequency rises and falls over time. This type of modulation is not observable when the timing information is reduced to the standard statistics. The DSP's can process the multiple blocks simultaneously and the multiple results are transferred to the tester in a single transfer. Test time can reduced by as much as 50% in some cases. Multiple Block mode can be illustrated by the case where a PLL's output frequency and RMS jitter are to be measured at three different VDD voltages. The conventional technique would be to repeat the measurement three different times and the total test time would be three times as long as a single VDD measurement. Multiple block mode can be used to capture the measurements of the three VDD levels in three sequential blocks, with only the VDD settling time between the captures. The three blocks are DSP processed in one operation and the results are transferred to the test in one transfer. Total test time is about 50% longer than for a single VDD. Femto PLL Shell Program The Guide Technology staff of applications engineers has developed a "Shell Program" to aid customers in developing their PLL test programs. The Shell Program is designed so that the user can get most of the standard PLL tests running quickly by simply enabling flags and setting variables in the definition file. The Shell Program was also designed to run on many different testers through the ATE-specific "User Function" capability. Benchmarks NOTE: These test times are based on Femto 2000 measurements taken in March 2002. Major throughput enhancements have been implemented since that improve these times by a factor of two or more.

Benchmark #1

Conditions: 8 parallel PLL channels 64 samples per measurement Test suite performed: VOL/VOH search Frequency Average Single Period Average Single Period RMS Jitter Single Period Peak to Peak Jitter Pulse Width Pulse Width RMS Jitter Rise Time Fall Time Approximate Test Time: 330 milliseconds

Benchmark #2

Conditions: 2 parallel PLL channels 400 samples Tests run at 3 VDD levels Test suite performed: Frequency Average Single Period RMS Jitter Approximate Test Time: 150 milliseconds.

Benchmark #3

Conditions: 2 parallel PLL channels Test suite performed: PLL Lock Time Measurement Results: Measured PLL 1 Lock Time: 40 microseconds Measured PLL 2 Lock Time: 45 microseconds Approximate Test Time: 45 milliseconds

Benchmark #4 Conditions: 2 parallel PLL channels Test suite performed: PLL Loop Bandwidth Measurement Results: Measured PLL 1 Loop Bandwidth: 72 KHz Measured PLL 2 Loop Bandwidth: 45 KHz Approximate Test Time: 40 milliseconds

 
 